Determination of Lead in Compound Fertilizer by AA-1800 Atomic Absorption Spectrometer
Key words: atomic absorption spectrometer; compound fertilizer;
;
This standard specifies the determination of lead in compound fertilizers by atomic absorption spectrometry.
Atomic absorption spectrometry
1
Subject content and scope of applicationThis standard specifies the determination of lead in compound fertilizers by atomic absorption spectrometry.
This standard applies to the determination of lead content is less than
25μg/mL
Sample solution.2
Reference StandardGB 602
Preparation of standard solution for determination of chemical reagent impuritiesGB 6682
Analytical laboratory water specifications and test methods3
Method summaryLead in the sample solution, in the air
-
Atomization in the acetylene flame, the resulting atomic vapor, absorbs the characteristic wavelength emitted from the lead hollow cathode lamp283.3nm
The amount of light absorbance is proportional to the atomic concentration of lead in the flame.4
Reagents and materialsAnalytically pure reagents are used in the analysis unless otherwise stated. The water used should be consistent with
GB 6682
Medium and third grade water specifications.4.1
hydrochloric acid(GB 622)
Solution:c(HCl)=0.5mol/L
. measure42mL
hydrochloric acid(p1.19)
Dilute to1L
Mix well;4.2
Lead standard solution:0.1mg/mL
. pressGB 602
in(4.82)
Formulated, this solution1mL
Lead100μg
;4.3
Dissolved acetylene(GB 6819)
, or the same specifications of acetylene.5
equipmentGeneral laboratory instruments, equipment and:
5.1 AA-1800C
Atomic absorption spectrophotometer: equipped with air-
Acetylene burner and lead hollow cathode lamp.6
Analysis step6.1
Drawing of the working curveBy table
1
As shown, the lead standard solution is placed separately6
One100mL
Hydrochloric acid solution in volumetric flask(4.1)
Dilute to the mark and mix.table
1
Lead standard solution
(4.2)
volume,mL
Corresponding lead content,
Gg/mL
0
5.0
10.0
15.0
20.0
25.0
0
5.0
10.0
15.0
20.0
25.0
Note: The preparation of the standard series solution can be adjusted according to the amount of lead in the sample and the sensitivity of the instrument.
Before performing spectrophotometry, according to the nature of the element to be tested, refer to the instrument manual for the spectral bandwidth, lamp current, burner height, air used for the measurement.
-
The acetylene flow ratio is selected for optimal working conditions.
Then, at the wavelength
283.3nm
Use air-
Acetylene oxide flame with lead content0
The standard solution is the reference solution, adjustedAtomic absorption spectrophotometer
After the absorbance was zero, the absorbance of each standard solution was measured. After each measurement, the burner was washed with a water spray.Lead content of each standard solution
(μg/mL)
For the abscissa, the corresponding absorbance is plotted on the ordinate and the working curve is drawn.6.2
DeterminationThe test solution is not diluted
(GB/T
14539.1
in5.2
Article)
Or pipetting a certain amount of test solution into a volumetric flask and using a hydrochloric acid solution(4.1)
Dilution as a sample solution for measurement(
The lead content of the solution to be tested must be less than25μg/mL)
Determine the absorbance of the test solution under the same conditions as the standard series solution, and find the corresponding lead concentration on the working curve.(μg/mL)
.7
Expression of analysis results7.1
When the test solution directly measures the absorbance without dilution, lead(Pb)
contentX5
, expressed in mass percent, by type(1)
Calculation:...................................................
(1)
In the formula:
c
─ the concentration of lead in the sample solution detected by the working curve,?g/mL
;m
─ the quality of the sample,g
;200
─ the total volume of the test solution,mL
.7.2
After taking a certain amount of test solution and diluting it to a certain volume, when measuring the absorbance, lead (Pb
)contentX5
, expressed in mass percent, by formula (2
) Calculation:................................................
(2)
In the formula:
c
─ the concentration of lead in the sample solution detected by the working curve,Gg/mL
;m
─ the quality of the sample,g
;200
─ the total volume of the test solution,mL
;V1
─ draw a certain amount of test solution,mL
;V2
─ the total volume of the sample solution,mL
;8
Allowable differenceTaking the arithmetic mean of the parallel determination results as the measurement result;
The relative deviation of the parallel determination results should be consistent with the table.
2
Claim:table
2
Lead content, %
Allow relative deviation, %
0.005
~0.0020
<
0.0020
~0.0001
<
0.0001
25
50
100
Our 5 wire resistive Touch Screen is the highest accuracy ,stable ,no effect to touch when the liquid pouring or spilling.And it is widely use to Industrial,Retail,
Transportation ,Finance,Hospital or other area.
5 wire touch screen have the following advantages,we worthy of your trust.
1) Can withstand more than 5,000,000 points.
2) 4096x4096 Resolution.
3) USB or RS232 Interface.
4) High accuracy and sensitivity.
5) Glass + Film Material.
6) Finger and touch pen activation.
7) Our products have passed CE, FCC, RoHS certification!
pictures show:
5 Wire Touch Screen Panel,Glass Touch Screen Panel,Touch Screen with Controller,Resistive Touch Screen,Waterproof Touch Screen,GPS Touch Screen
ShenZhen GreenTouch Technology Co.,Ltd , https://www.bbstouch.com